Visual field assessment in glaucoma: Comparative evaluation of manual Kinetic Goldmann Perimetry and Automated Static Perimetry

oleh: Agarwal Harish, Gulati V., Sihota Ramanjit

Format: Article
Diterbitkan: Wolters Kluwer Medknow Publications 2000-01-01

Deskripsi

<b>Purpose:</b> To compare the detection and assessment of progression of visual field defects in primary open-angle glaucoma with manual suprathreshold perimetry on Goldmann perimeter and automated static threshold perimetry on Humphery visual field (HVF) analyzer. <b>Methods:</b> 105 eyes of 54 patients of primary open-angle glaucoma were followed up with 3-monthly perimetry on Goldmann perimeter and HVF analyzer, for a period of 9 months. <b>Results:</b> HVF analyzer picked up visual field defects in 48 (46&#x0025;) eyes whereas Goldmann perimeter picked up visual field defects in 26 (25&#x0025;) eyes. HVF analyzer demonstrated progression in 14 eyes whereas Goldmann perimeter detected progression in 7 eyes during follow up of 9 months. <b>Conclusions:</b> HVF analyzer is superior to Goldmann perimeter to document and to demonstrate progression of visual field defects in primary open-angle glaucoma.