Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems

oleh: Anastasios Tentolouris, Ioanna Eleftheriadou, Pinelopi Grigoropoulou, Maria Nikoloudi, Evangelia Siami, Diamantis I. Tsilimigras, Nikolaos Tentolouris

Format: Article
Diterbitkan: JCDR Research and Publications Private Limited 2018-06-01

Deskripsi

Introduction: Self-Monitoring of Blood Glucose (SMBG) is essential for achieving glycaemic control in individuals with diabetes mellitus. Aim: The aim of the study was to evaluate the accuracy and the usability of six different commercially available SMBG systems. Materials and Methods: A total of 120 individuals with both types of diabetes, aged between 18-80 years, were recruited. Capillary Blood Glucose (BG) was measured on six SMBG systems, while the YSI 2300 STAT PLUS was used as a reference method. Accuracy performance was based on requirements resembling the ISO 15197:2013 (95% of the BG results within ±15 mg/dL of the reference method at BG concentrations <100 mg/dL and within ±15% of the reference method at BG concentration ≥100 mg/dL). Usability testing of the SMBG was performed using the Likert Scale Questionnaire. Results: Four out of six and five out of six glucometers demonstrated excellent performance in BG values <100 mg/dL (n=40) and in BG values ≥100 mg/dL (n=80), respectively. Across the overall BG range tested, five out of six SMBG systems had good accuracy performance. Among the three SMBG systems with the highest performance (>99%) in the overall BG range (29-475 mg/dL) the one with an innovative strip insertion technique was less favoured by the participants. Conclusion: Most of the examined commercially available SMBG devices demonstrated adequate performance when they were tested in requirements resembling the ISO 15197:2013. Regarding usability, the Accu-Chek Aviva Plus and the Contour Next were the most convenient devices.