Find in Library
Search millions of books, articles, and more
Indexed Open Access Databases
Chip-Oriented Fluorimeter Design and Detection System Development for DNA Quantification in Nano-Liter Volumes
oleh: Da-Sheng Lee, Ming-Hui Chen
Format: | Article |
---|---|
Diterbitkan: | MDPI AG 2009-12-01 |
Deskripsi
The chip-based polymerase chain reaction (PCR) system has been developed in recent years to achieve DNA quantification. Using a microstructure and miniature chip, the volume consumption for a PCR can be reduced to a nano-liter. With high speed cycling and a low reaction volume, the time consumption of one PCR cycle performed on a chip can be reduced. However, most of the presented prototypes employ commercial fluorimeters which are not optimized for fluorescence detection of such a small quantity sample. This limits the performance of DNA quantification, especially low experiment reproducibility. This study discusses the concept of a chip-oriented fluorimeter design. Using the analytical model, the current study analyzes the sensitivity and dynamic range of the fluorimeter to fit the requirements for detecting fluorescence in nano-liter volumes. Through the optimized processes, a real-time PCR on a chip system with only one nano-liter volume test sample is as sensitive as the commercial real-time PCR machine using the sample with twenty micro-liter volumes. The signal to noise (S/N) ratio of a chip system for DNA quantification with hepatitis B virus (HBV) plasmid samples is 3 dB higher. DNA quantification by the miniature chip shows higher reproducibility compared to the commercial machine with respect to samples of initial concentrations from 103 to 105 copies per reaction.