Eccentric mitral regurgitation through the LAmbre closure device

oleh: Manuel Barreiro-Pérez, María Elena Tundidor-Sanz, Ignacio Cruz-González

Format: Article
Diterbitkan: Permanyer 2019-08-01

Deskripsi

An 83-year-old female with severe asymptomatic mitral regurgitation (MR) was admitted for an elective percutaneous left atrial appendage occlusion (LAAO) due to severe gastrointestinal bleeding under anticoagulant treatment. The transesophageal echocardiography (TEE) conducted before the procedure confirmed the presence of severe MR due to a flail P2 segment with a jet lesion on the sliding atrial wall adjacent to the appendage. An LAAO procedure was performed and the LAmbre system (Lifetech Scientific) was successfully deployed. A first assessment confirmed the presence of MR jet over the external lobe of the device (figure 1A; LA, left atrium; LV, left ventricle), but detailed scanning detected one MR jet pathway below the LAAO device (figure 1B,C). The 3D-TEE of the mitral valve P2 scallop (*) is shown on figure 1D-F (Ao, aortic root; LAAO device (#); the 3D-volume is cropped according to red line to create figure E). The pre-LAmbre device fluoroscopy image is shown on figure G. The steps of this procedure are available at video 1 of the supplementary data. Device change or relocation were not considered. We thought that a LAmbre device was the best option for a cone-shaped appendage since the external disc is far enough from the mitral...