Find in Library
Search millions of books, articles, and more
Indexed Open Access Databases
Effect of lens constants optimization on the accuracy of intraocular lens power calculation formulas for highly myopic eyes
oleh: Jia-Qing Zhang, Xu-Yuan Zou, Dan-Ying Zheng, Wei-Rong Chen, Ao Sun, Li-Xia Luo
Format: | Article |
---|---|
Diterbitkan: | Press of International Journal of Ophthalmology (IJO PRESS) 2019-06-01 |
Deskripsi
AIM: To evaluate the effect of different lens constant optimization methods on the accuracy of intraocular lens (IOL) power calculation formulas for highly myopic eyes. METHODS: This study comprised 108 eyes of 94 consecutive patients with axial length (AL) over 26 mm undergoing phacoemulsification and implantation of a Rayner (Hove, UK) 920H IOL. Formulas were evaluated using the following lens constants: manufacturer’s lens constant, User Group for Laser Interference Biometry (ULIB) constant, and optimized constant for long eyes. Results were compared with Barrett Universal II formula, original Wang-Koch AL adjustment method, and modified Wang-Koch AL adjustment method. The outcomes assessed were mean absolute error (MAE) and percentage of eyes with IOL prediction errors within ±0.25, ±0.50, and ±1.0 diopter (D). The nonparametric method, Friedman test, was used to compare MAE performance among constants. RESULTS: Optimized constants could significantly reduce the MAE of SRK/T, Hoffer Q, and Holladay 1 formulas compared with manufacturer’s lens constant, whereas the percentage of eyes with IOL prediction errors within ±0.25, ±0.50, and ±1.0 D had no statistically significant differences. Optimized lens constant for long eyes alone showed non-significant refractive advantages over the ULIB constant. Barrett Universal II formula and formulas with AL adjustment showed significantly higher accuracy in highly myopic eyes (P<0.001). CONCLUSION: Lens constant optimization for the subset of long eyes reduces the refractive error only to a limited extent for highly myopic eyes.