Evaluation of left ventricular ejection fraction by a new automatic tool on a pocket ultrasound device: Concordance study with cardiac magnetic resonance imaging.

oleh: Lucie Berger, Fabien Coisy, Skander Sammoud, Fabien de Oliveira, Romain Genre Grandpierre, Laura Grau-Mercier, Xavier Bobbia, Thibaut Markarian

Format: Article
Diterbitkan: Public Library of Science (PLoS) 2024-01-01

Deskripsi

<h4>Introduction</h4>Assessment of left ventricular ejection fraction (LVEF) is one of the primary objectives of echocardiography. The gold standard assessment technique in emergency medicine is eyeballing. A new tool is now available on pocket ultrasound devices (PUD): automatic LVEF. The primary aim of this study was to evaluate the concordance between LVEF values estimated by automatic LVEF with PUD and by cardiac magnetic resonance imaging (MRI).<h4>Materials</h4>This was a prospective, monocentric, and observational study. All adult patients with an indication for cardiac MRI underwent a point-of-care ultrasound. Blinded to the MRI results, the emergency physician assessed LVEF using the automatic PUD tool and by visual evaluation.<h4>Results</h4>Sixty patients were included and analyzed. Visual estimation of LVEF was feasible for all patients and automatic evaluation for 52 (87%) patients. Lin's concordance correlation coefficient between automatic ejection fraction with PUD and by cardiac MRI was 0.23 (95% CI, 0.03-0.40).<h4>Conclusion</h4>Concordance between LVEF estimated by the automatic ejection fraction with PUD and LVEF estimated by MRI was non-existent.